The Feasibility and Safety of the Glean Urodynamics System Study.

  • May 1, 2025
  • 1 min read

Updated: Mar 30


Thirty-eight participants were enrolled. Glean sensor insertion was attempted in 33 participants and successful in 32 (97.0%). The median sensor insertion time was 33.62 seconds (range: 12.32–256). The Glean Urodynamics System recorded vesical pressure, and all 32 participants were able to void with the sensor indwelling. Median sensor removal time was 5.71 seconds (range: 0.79–22.15). There were 14 adverse events in 12 participants, of which 7 events (in 7 participants) were attributed to the Glean Urodynamics System. There were no serious adverse events. Overall participant feedback on the Glean Urodynamics System was positive, and clinicians reported it was easy to use.




Comments


bottom of page